Answers are grounded in this paper and cite their sources.
A versatile irradiator has been constructed for in vitro irradiation of mammalian cells with alpha-particles of well-defined energy, LET, direction, dose and dose rate. It is based on approximately 1.2 x 10(9) Bq of 238Pu (on a platinum disc) contained in a He-filled chamber. In a standard configuration, monolayers of cells grown in 10 Hostaphan-based dishes are irradiated with 3.26 +/- 0.22 MeV alpha-particles (LET 121 keV microns-1) at selectable dose rates from approximately 2 Gy min-1 down to less than 10(-4) Gy min-1 (i.e. fluence rates of 1 x 10(7) cm-2 min-1 to 3 x 10(2) cm-2 min-1). Single dishes can be irradiated at dose rates up to 24 Gy min-1 (fluence rate 1 x 10(8) cm-2 min-1). Incident energy and LET can be varied from 0.8 to 4.2 MeV and 266 to 102 keV microns-1, respectively. The irradiator has full incubation and gassing facilities for protracted irradiations. The irradiator is particularly suitable for in vitro analytical studies of the biological effects of alpha-particles of energies and LETs similar to those which cells may receive in vivo from radionuclides such as radon and the actinides. It has been used successfully for investigations of a variety of alpha-particle-induced effects in different cell types irradiated either as attached monolayers or as very thin suspensions.
No takes yet. Share an insight, caveat, or question.
Goodhead et al. (1991) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: