Key result
Surgical staplers and double-layered sutures were superior to single-layered sutures in resilience to pneumatic pressure.
Why the study?
Does the use of surgical staplers or double-layered sutures improve resilience to pneumatic pressure compared to single-layered sutures for epicardial LAA closure in an ex vivo model?
Does the use of surgical staplers or double-layered sutures improve resilience to pneumatic pressure compared to single-layered sutures for epicardial LAA closure in an ex vivo model?
Answers are grounded in this paper and cite their sources.
In an ex vivo porcine model, surgical staplers and double-layered sutures provided significantly higher resilience to pneumatic pressure for LAA closure compared to single-layered sutures or purse strings.
BACKGROUND: Formation of clots in the left atrial appendage (LAA) may lead to embolism and consecutive cerebral stroke. This risk is reduced by closure and resection. To address the efficacy of surgical LAA closure, resilience to pneumatic pressure was studied. Different surgical techniques were compared in an experimental model. METHODS: From freshly slaughtered pigs cardiopulmonary preparations were taken. The left atrium was clamped airtight and the LAA was cannulated. Via a manually operated pump pressure was applied and a digital pressure gauge was connected. Four groups (each n=12) with different epicardial closures were studied: (I) purse string; (II) single layered continuous suture; (III) double layered suture; (IV) surgical stapler. A nonparametric test was used for group comparisons of mean burst pressures (mmHg). Statistical significance was defined at P<0.05. RESULTS: Mean burst pressures in group 1 amounted to 97.1±13.0 mmHg, in group 2 to 105.5±13.3 mmHg, in group 3 to 124.6±14.2 mmHg and in group 4 to 136.6±12.5 mmHg. Excepting differences between groups 1 and 2 comparisons between groups were significantly different. CONCLUSIONS: model surgical staplers and double layered hand crafted sutures proved well suitable for closure of the LAA. They were significantly superior to single layered sutures in terms of resilience to pneumatic pressure. This may be relevant to durability and should be discussed with regard to clinical choice.
No takes yet. Share an insight, caveat, or question.
Mirow et al. (2017) studied Left atrial appendage closure (n=48). Surgical stapler and double layered suture vs. Purse string and single layered continuous suture was evaluated on Mean burst pressure (mmHg) (p=<0.05). Surgical staplers and double-layered sutures were superior to single-layered sutures in resilience to pneumatic pressure.
Synapse has enriched 10 related papers. Consider them for comparative context: