Answers are grounded in this paper and cite their sources.
In magnetic resonance imaging, time-varied gradient magnetic fields may stimulate nerves and muscles by inducing electric fields and currents in patients, which may potentially cause health problem. In this study, realistic transversex-,y-gradient coils and longitudinalz-gradient coil were numerically designed. The magnetic flux density, current density and electric fields in a 3-D real human man model were calculated by employing impedance method. The exposure levels have been compared with the guidelines of the International Commission on Non-Ionizing Radiation Protection (ICNIRP). It was found that the spatially averaged magnetic flux density greatly exceeds the ICNIRP reference level, and the maximum value of the 1-cm2-averaged induced current density in central nervous system greatly exceeds the ICNIRP basic restriction. The strongest levels of exposure are observed when all three gradient coils are operated simultaneously.
No takes yet. Share an insight, caveat, or question.
Lu et al. (2011) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: