Key result
The SensAheart point-of-care assay demonstrated similar overall diagnostic accuracy to hs-cTnT for the early diagnosis of acute coronary syndrome (82.7% vs 76.0%, p=0.08).
Why the study?
Early and accurate diagnosis of ACS is essential for initiating lifesaving interventions, prompting evaluation of the diagnostic performance of a novel point-of-care rapid assay.
Does the SensAheart point-of-care assay provide comparable diagnostic accuracy to hs-cTnT for acute coronary syndrome in patients presenting with typical chest pain?
Observational (n=225)
Does the SensAheart point-of-care assay provide comparable diagnostic accuracy to hs-cTnT for acute coronary syndrome in patients presenting with typical chest pain?
Absolute Event Rate: 82.7% vs 76%
p-value: p=0.08
The SensAheart point-of-care assay provides similar overall diagnostic accuracy to hs-cTnT for early detection of ACS, with higher sensitivity but lower specificity.
BACKGROUND: Early and accurate diagnosis of acute coronary syndrome (ACS) is essential for initiating lifesaving interventions. In this article, the diagnostic performance of a novel point-of-care rapid assay (SensAheart©) is analyzed. This assay qualitatively determines the presence of 2 cardiac biomarkers troponin I and heart-type fatty acid-binding protein that are present soon after onset of myocardial injury. METHODS: We conducted a prospective observational study of consecutive patients who presented to the emergency department with typical chest pain. Simultaneous high-sensitive cardiac troponin T (hs-cTnT) and SensAheart testing was performed upon hospital admission. Diagnostic accuracy was computed using SensAheart or hs-cTnT levels versus the final diagnosis defined as positive/negative. RESULTS: Of 225 patients analyzed, a final diagnosis of ACS was established in 138 patients, 87 individuals diagnosed with nonischemic chest pain. In the overall population, as compared to hs-cTnT, the sensitivity of the initial SensAheart assay was significantly higher (80.4 vs. 63.8%, p = 0.002) whereas specificity was lower (78.6 vs. 95.4%, p = 0.036). The overall diagnostic accuracy of SensAheart assay was similar to the hs-cTnT (82.7% compared to 76.0%, p = 0.08). CONCLUSIONS: Upon first medical contact, the novel point-of-care rapid SensAheart assay shows a diagnostic performance similar to hs-cTnT. The combination of 2 cardiac biomarkers in the same kit allows for very early detection of myocardial damage. The SensAheart assay is a reliable and practical tool for ruling-in the diagnosis of ACS.
No takes yet. Share an insight, caveat, or question.
Alcalai et al. (2020) conducted an observational in Acute coronary syndrome (n=225). SensAheart assay vs. High-sensitive cardiac troponin T (hs-cTnT) was evaluated on Overall diagnostic accuracy for acute coronary syndrome (p=0.08). The SensAheart point-of-care assay demonstrated similar overall diagnostic accuracy to hs-cTnT for the early diagnosis of acute coronary syndrome (82.7% vs 76.0%, p=0.08).