Answers are grounded in this paper and cite their sources.
A room temperature phosphorimetric (RTP) procedure was used for the determination of 6-thioguanine (6-TG). The method is based on paper substrate room temperature phosphorimetry (PS-RTP) using indium sulfate, In2(SO4)3 as a heavy atom perturber. Various factors affecting the room temperature phosphorescence of 6-TG are discussed. The linear dynamic range for 6-TG is from 3.3 to 334.3 ng per spot with a detection limit of 4.6 ng per spot and a relative standard deviation (RSD) of 2.38%. The recovery of standard 6-TG added to commercial tablets is in the range 96.39-98.44%. The method is simple, rapid and sensitive and can be applied to the analysis of commercial tablets without interference.
No takes yet. Share an insight, caveat, or question.
Yang et al. (2000) studied this question.