Key result
Fluorescent LDL demonstrates ~2-fold higher lymphocyte LDL receptor affinity than unlabeled LDL.
Answers are grounded in this paper and cite their sources.
We have developed a simple method for comparing the relative fluorescence intensity (FI) of flow cytometry histograms. It entails assessment of the FI (equivalent to the fluorescence-activated cell sorter (FACS) channel) of the 50th or 75th percentiles of either positively stained cells or the total cell population. We illustrate the method with dilution curves of 1) monoclonal antibodies against the T4 surface antigen of human peripheral blood lymphocytes and 2) fluorescent low density lipoprotein (LDL) binding to the human peripheral blood lymphocytes LDL receptor. We demonstrate the versatility of the method by characterizing the binding properties of fluorescent LDL to their receptors. Binding was shown to be specific and of high affinity, and to reach a steady state plateau at about 2 hr; the affinity of fluorescent LDL for the receptor was found to be two to three times higher than that of the unlabeled LDL.
No takes yet. Share an insight, caveat, or question.
Traill et al. (1986) studied this question. 50th or 75th percentile relative fluorescence intensity flow cytometry method (using monoclonal antibodies against T4 and fluorescent LDL) vs. Unlabeled LDL was evaluated on Specificity, affinity, and time to steady-state plateau of fluorescent LDL binding to human peripheral blood lymphocyte LDL receptors. Fluorescent LDL binding to lymphocyte LDL receptors analyzed via 50th/75th percentile flow cytometry reached steady state at ~2 hours with 2- to 3-fold higher affinity than unlabeled LDL.
Synapse has enriched 10 related papers. Consider them for comparative context: