Why the study?
The comparative effectiveness and safety of left atrial appendage closure (LAAC) devices, compared with each other and with anticoagulation, is unclear.
Does left atrial appendage closure with the Amplatzer Amulet device reduce the risk of stroke and all-cause death with similar safety compared to the Watchman device and oral anticoagulation in patients with atrial fibrillation?
Does left atrial appendage closure with the Amplatzer Amulet device reduce the risk of stroke and all-cause death with similar safety compared to the Watchman device and oral anticoagulation in patients with atrial fibrillation?
This network meta-analysis of 6 RCTs demonstrates that the Amplatzer Amulet and Watchman left atrial appendage closure devices have comparable effectiveness for preventing stroke and all-cause mortality, though the Amulet device is associated with a two-fold higher risk of pericardial effusion.
Answers are grounded in this paper and cite their sources.
Supports individualized LAAC device selection by anatomy or operator expertise; extends RCT evidence for device equivalence versus anticoagulation in AF.
Introduction Atrial fibrillation‐related stroke is a leading cause of morbidity and mortality. The comparative effectiveness and safety of left atrial appendage closure (LAAC) devices, compared with one another and with anticoagulation, is unclear. Methods We conducted a systematic review and network meta‐analysis (NMA) of all clinical trials comparing the Watchman and Amplatzer Amulet LAAC devices with each other or with warfarin or direct oral anticoagulants (DOACs). The primary comparison was between LAAC devices with secondary comparisons to anticoagulation. The primary effectiveness outcomes were any stroke and all‐cause death. Safety outcomes included any thromboembolism, device embolization, and pericardial effusion. Results There were 476 articles identified from the search and 6 eligible RCTs were included ( n = 3666). There was no difference in the risk of stroke with Amulet versus Watchman (RR = 1.48, 95% CI: 0.64–3.46, I 2 = 41.3%), nor in the risk of death (RR = 1.00, 95% CI: 0.59–1.70, I 2 = 45.0%). Risk of thromboembolism was not significantly different with Amulet versus Watchman (RR = 0.73, 95% CI: 0.18–2.97, I 2 = 0%), nor was risk of device embolization (RR = 2.29, 95% CI: 0.71, 7.43, I 2 = 0%). Both devices exhibited increased risk of pericardial effusion compared with warfarin, with Amulet at highest relative risk (RR = 27.08, 95% CI: 3.53–207.98, I 2 = 0%) followed by Watchman (RR = 12.79, 95% CI: 1.73–94.85, I 2 = 0%). Amulet also carried higher relative risk of pericardial effusion than Watchman (RR = 2.12, 95% CI: 1.45–3.09). Conclusion In this NMA, the Amulet and Watchman LAAC devices were associated with similar risks for stroke, mortality, thromboembolism, and device embolization. Pericardial effusion risk was higher with Amulet than Watchman.
No takes yet. Share an insight, caveat, or question.
Davis et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: