The contrast formation and the achievable lateral resolution of a scanning capacitance microscope have been simulated using the finite element method. On conducting surfaces, a resolution of about 2 nm is expected with probes with about 5–10 nm radius of curvature. At smaller radii, the resolution degrades due to the decrease of the local contribution to the integral capacitance sensed by the probe. The lateral resolution is less affected by the radius of the tip than by the tip-to-conducting base distance, though sharper tips produce higher contrast. With a tip radius of approximately 25 nm, resolution of features 5 nm in diameter has been achieved on gold-coated silicon.