The performance of the average infra-red spectrometer is conditioned mainly by the characteristics of the detector employed. Assuming a minimum usable energy level at the exit slit, the square of the minimum frequency band embraced by the exit slit is proportional to the minimum energy level and to the focal length of the collimator, and inversely proportional to the angular dispersion of the prism or grating, the area of the collimator mirror, the energy flux at the entrance slit and the slit length. If, however, the time of response of the detector is taken into account, the inter-relation of spectrum scanning speed, signal-to-noise ratio and resolving power must be considered. For this purpose a simplified model of an infra-red detector has been used, defined by three parameters. This allows the relative merits of detectors to be assessed and the performance of any given spectrometer with a specified detector to be estimated. The inter-relation of resolving power and spectral scanning speed has been examined for "chopped-beam" and "video" presentation.