Answers are grounded in this paper and cite their sources.
The versatility of the quadrupole ion trap as a mass spectrometer in which the mass resolution may be varied over a wide range has been demonstrated. Mass resolution of 0.8 × 10 7 and 1.2 × 10 7 have been obtained for ions of m / z 414 and m / z 614, respectively, well in excess of the normal value of 3 m , where m is the mass of ion. At the highest mass resolution, peak‐widths at half‐maximum are but 52 × 10 −6 u. While it has been shown previously that enhanced mass reslution in the quadrupole ion trap can be achieved by reduction of the mass scanning rate, the range of mass resolutions reported here were achieved by reducing the mass scanning rate, in stages, by an overall factor of ca 5.5 × 10 4 .
No takes yet. Share an insight, caveat, or question.
Londry et al. (1993) studied this question.