Answers are grounded in this paper and cite their sources.
The Sensititre Automated Reading and Incubation System (ARIS) performs automated reading of Sensititre identification and susceptibility plates. ARIS susceptibility results were compared to manually read results. A total of 212 isolates (3352 organism-antimicrobial agent combinations) yielded 99.4% essential agreement and categorical error rates of 0.2% minor, 0% major, and 1.0% very major. ARIS yields accurate susceptibility results compared to plates read manually.
No takes yet. Share an insight, caveat, or question.
Chapin et al. (2004) studied this question.