Why the study?
Is it feasible to place a Nitinol inferior vena cava filter guided by high-resolution MR fluoroscopy in a pig model?
Is it feasible to place a Nitinol inferior vena cava filter guided by high-resolution MR fluoroscopy in a pig model?
MR-guided placement of a Nitinol inferior vena cava filter is feasible in a pig model, allowing for non-contrast evaluation using 3D phase contrast MRA.
Answers are grounded in this paper and cite their sources.
Feasibility in porcine model supports preclinical development; leaves open translation to human IVC filter placement.
Percutaneous placement of an inferior vena cava filter is a means for long-term prevention of pulmonary thromboembolism. In this study we investigated the magnetic resonance (MR) imaging properties of a Nitinol vena cava filter, in various anatomic and angiographic scans, as well as the feasibility of placing this filter under near real-time, high-resolution MR fluoroscopy. We made use of the passive tracking strategy, with on-line image processing and visualization, both in vitro and in a pig. The artifacts provoked by the metallic filter were such that the position and orientation of the filter were well depicted in all scans. Considerable radiofrequency caging obscured the interior of the filter. Our experiments showed that an MR-guided vena cava filter placement, with sufficient temporal and spatial resolution, is possible. Three-dimensional phase contrast MRA allowed direct evaluation of the filter placement procedure, without the use of contrast agent.
No takes yet. Share an insight, caveat, or question.
Bartels et al. (2000) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: