The Raman spectrum of the organophosphorus pesticide fenthion is presented, and a novel Fourier transform (FT)-Raman method for the quantitative determination of fenthion in pesticide formulations is proposed. Calibration curves based on Raman intensity and band area at 2951, 1065, 661, and 604 cm −1 were linear (correlation coefficients: 0.991–0.998) in the concentration range of 0.4–4.36 M. Detection limits ranged between 0.14 to 0.36 M and precision from 0.4 to 6.8% relative standard deviation (RSD) ( n = 4). Spectra normalization against the 802 cm −1 cyclohexane band proved to be effective in minimizing the effect of excitation intensity fluctuations and sample positioning changes in calibration data, resulting in improved long-term stability of the calibration curves. Results obtained during the analysis of the Lebaycid formulation compare well with those obtained by the gas chromatography (GC) reference method. The proposed FT-Raman method is fast, simple, and safe, avoids sample pretreatment, and minimizes handling of toxic samples.
No takes yet. Share an insight, caveat, or question.
Skoulika et al. (1999) studied this question.