Answers are grounded in this paper and cite their sources.
OBJECTIVE: To validate a transit time ultrasound flowmeter (CardioMed CM 4000) for measuring blood flow in isolated colonic mesenteric arteries. DESIGN: Experimental and clinical study. SETTING: Teaching hospital, Denmark. ANIMALS AND SUBJECTS: One female pig, and 6 patients being operated on for carcinoma of the sigmoid colon and rectum. INTERVENTIONS: Volume blood flow measured by Cardiomed CM 4000 and pump withdrawal flow recordings. MAIN OUTCOME MEASURES: Correlation between the two methods. RESULTS: There was good agreement between transit time flow recordings and pump withdrawal flow recordings (correlation coefficient of 1.0). Of the differences between the two methods, 95% were between -0.16 ml min(-1) and 1.29 ml min(-1), mean 0.57 ml min(-1), or (in percentages) 105, 95% lying between 97-115. There was also good reproducibility in transit time flow recordings, the mean difference between repeated measurements being 0.06 ml min(-1), 95% lying between -0.66 ml min(-1) and 0.78 ml min(-1). CONCLUSION: Ultrasound transit time flow recordings gave precise measurements of blood flow in isolated colonic mesenteric arteries.
No takes yet. Share an insight, caveat, or question.
Mortensen et al. (2002) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: