Does pulsed doppler ultrasound provide comparable cardiac output measurements to dye dilution in children ventilated after cardiac surgery?
Does pulsed doppler ultrasound provide comparable cardiac output measurements to dye dilution in children ventilated after cardiac surgery?
Pulsed doppler ultrasound is a reliable, noninvasive alternative to dye dilution for measuring cardiac output in ventilated children after cardiac surgery.
Answers are grounded in this paper and cite their sources.
May support pulsed Doppler as noninvasive cardiac output option in ventilated post-cardiac surgery children; leaves open prospective validation before routine use.
Dye dilution was compared with pulsed doppler for the measurement of cardiac output in eighteen children being ventilated after cardiac surgery. The mean difference between the two techniques was -0.04/min (dye minus doppler) with 95% confidence limits of 0.25 l/min and -0.33 l/min over a cardiac output range 0.27-6.12 l/min; this difference is not significant. Calculation of the product-moment correlation coefficient showed a close relationship between the dye dilution and doppler methods with r = 0.97. Pulsed doppler is a new noninvasive technique that can be used instead of dye dilution for the measurement of cardiac output in co-operative or anaesthetised children.
No takes yet. Share an insight, caveat, or question.
Tibballs et al. (1988) studied this question.