This is the authors' abstract. We don't add key points for this paper.
The two-modulator generalized ellipsometer has been used to measure samples in transmission. In this configuration, the instrument can completely characterize a linear diattenuator and retarder, measuring birefringence, diattenuation, the angle of the principal axis, and the sample depolarization simultaneously and accurately. This instrument can be operated in two modes: (1) spectroscopic, in which measurements are made through the entire sample aperture as a function of wavelength, and (2) spatially resolved, in which measurements are made at a single wavelength and a birefringence picture is made of the sample. Current spatially resolved measurements have been made at a resolution of approximately 40 microm. Four samples have been examined with this instrument: (1) a mica plate, (2) a Polaroid polarizer, and (3) two quartz plates.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Jellison et al. (2002) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: