Key result
BEV yields similar invasive gradients to SEV in failing bioprostheses, despite higher echo-derived gradients.
Why the study?
The hemodynamics of TAVI in degenerated transcatheter or surgical aortic valves (TAV-in-TAV or TAV-in-SAV) were unknown.
Does balloon-expandable TAV compared to self-expanding TAV result in different transaortic pressure gradients in patients with failing aortic bioprostheses?
Cohort (n=56)
No
Does balloon-expandable TAV compared to self-expanding TAV result in different transaortic pressure gradients in patients with failing aortic bioprostheses?
Absolute Event Rate: 6% vs 7%
p-value: p=0.109
In patients undergoing TAV-in-TAV or TAV-in-SAV, balloon-expandable valves show higher echocardiographic gradients but similar invasive gradients compared to self-expanding valves, highlighting a significant measurement discordance.
Discordance warrants caution in relying on echocardiography alone for valve assessment; leaves open standardization of post-TAVI gradients in valve-in-valve procedures.
Background Transcatheter aortic valve implantation (TAVI) is an established treatment for a failing aortic bioprosthesis. The hemodynamics of TAVI in degenerated transcatheter or surgical aortic valves (TAV-in-TAV or TAV-in-SAV) are unknown. We aimed to investigate hemodynamic differences between self- and balloon-expandable TAV-in-TAV and TAV-in-SAV groups, and transthoracic echocardiography-derived (TTE) versus invasive transaortic pressure gradients. Methods Patients ≥ 18 years with a self-expanding EVOLUT (SEV) or balloon-expandable SAPIEN3 (BEV) valve for TAV-in-TAV or TAV-in-SAV were included. Transaortic gradients were determined invasively and by TTE within 48 h post-intervention. Results We identified 56 patients with SEV ( n = 36) or BEV ( n = 20) TAV in a failing aortic bioprosthesis. Fourteen cases involved failing transcatheter valves and 42 surgical bioprostheses. Invasive mean gradients were similar after BEV and SEV (median (25th–75th percentile): 6.0 (1.5–6.5) vs 7.0 (2.0–10.0) mm Hg, p = 0.109). Mean gradients by TTE were higher for BEV than SEV at discharge (16.0 (10.8–19.8) versus 10.0 (7.0–12.0) mm Hg, p = 0.003) and 12-month follow-up (13.0 (11.0–15.5) versus 9.0 (7.0–12.3) mm Hg, p = 0.025). The discrepancy between TTE and invasive mean gradients was significantly larger for BEV than SEV (13.0 (6.3–14.0) vs 3.0 (1.5–8.3) mm Hg, p = 0.001) and most pronounced in TAV-in-SAV (BEV 13.0 (6.3–17.0) vs SEV 3.0 (1.0–8.5) mm Hg, p = 0.004). Conclusion Mean gradient after TAV in a failing bioprosthesis is similar for BEV and SEV, when measured invasively, but higher with BEV than SEV by TTE. The discordance between invasive and TTE-derived mean gradients is the largest in BEV.
No takes yet. Share an insight, caveat, or question.
Chatterjee et al. (2026) conducted a cohort in Failing aortic bioprosthesis (n=56). Balloon-expandable SAPIEN3 valve vs. Self-expanding EVOLUT valve was evaluated on Invasive mean transaortic pressure gradient post-intervention (p=0.109). In patients with failing aortic bioprostheses, invasive mean gradients were similar after balloon-expandable and self-expanding valve implantation (6.0 vs 7.0 mm Hg, p=0.109), but echocardiography-derived gradients were significantly higher with balloon-expandable valves.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: