This editorial highlights the engineering principles and clinical potential of using four-terminal intracardiac impedance to monitor stroke volume in patients with implantable devices.
This editorial refers to ‘Intracardiac impedance monitors stroke volume in resynchronization therapy patients’ by M. Bocchiardo et al. , on page 702. Nothing is more basic to life than cardiac output. However, an accurate method to monitor cardiac output with an implantable device has long eluded device developers. In this issue of the Journal, Bocchiardo et al. 1 present significant progress toward that end with encouraging human data. Electrical impedance is defined as the ratio of voltage (across) to current (through an object). This is typically measured by a two-terminal method. With the two-terminal method, the same contacts are used to inject the current and measure the voltage (or, equivalently, to impress the voltage and measure the current). This works quite well for most electronic components such as a discrete resistor found at an electronics store. However, in a fluid medium, the two-terminal method can introduce significant errors. Consider a glass aquarium filled with blood that is 20 cm long with a cross section of 10 × 10 cm (giving the cross-sectional area of 100 cm2). With the typical bulk resistivity of 150 Ωcm for blood the resistance, from end to end, of the blood will be: ![Formula][1] Now imagine that we attempt to measure the impedance using 2 mm diameter (1 mm radius) electrodes. The impedance from the electrode to the nearest 1 cm (10 mm) of fluid may be calculated by summing the resistance of the spherical shells which each have a resistance of: ![Formula][2] where ρ is the resistivity of the fluid (150 Ωcm or 1500 Ωmm for blood) and r … *Corresponding author. Tel: +1 805 428 1838; fax: +1 952 471 2482, Email: mark{at}kroll.name [1]: /embed/graphic-1.gif [2]: /embed/graphic-2.gif
No takes yet. Share an insight, caveat, or question.
Mark W. Kroll (2010) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: