At a site near Boulder, Colorado, simultaneous atmospheric measurements were made of NO, NO x , and NO y in a field intercomparison of instruments involving two currently employed techniques of NO x and NO y measurement. Both NO y instruments depended upon the reduction of the NO y species to NO with detection by chemiluminescence, but different catalysts were employed for the reduction: (1) a gold catalyst (with addition of 0.3% CO) at 300°C, or (2) a molybdenum catalyst at 400°C. The two systems of NO x detection involved (1) photolysis of NO 2 , and (2) reduction of NO 2 by solid ferrous sulfate. Several times during the intercomparison the response to calibrated samples of reference compounds NO, NO 2 , PAN, HNO 3 , n ‐propyl nitrate (NPN), and NH 3 was measured. From the results the following conclusions were made: (1) The two NO instruments agreed on NO mixing ratios that were measured during the daytime hours over a range from the limits of detection to 35 parts per billion by volume (ppbv), (2) the two NO y instruments gave similar estimates of NO y in ambient air over a wide range of mixing ratios (0.4–10 ppbv), and (3) the ferrous sulfate converter used for NO x detection showed a significant interference from NPN and PAN.