The Nellcor N-101 pulse oximeter provides accurate noninvasive oxygen saturation readings compared to arterial blood measurements in adult patients under anesthesia or intensive care.
The accuracy of the Nellcor N-101 pulse oximeter has been evaluated in adult patients receiving general anaesthesia or intensive care. Readings obtained noninvasively with this instrument were compared with measurements made on arterial blood using a Radiometer OSM2 oximeter. The pulse oximeter was easy to use and within the range tested (70-100 percent saturation of haemoglobin with oxygen) the readings were within I digit of the values obtained by in vitro measurement.
No takes yet. Share an insight, caveat, or question.
TYTLER et al. (1986) studied this question.