A pair of secondary standard tungsten strip lamps have had a luminance temperature – current calibration, in the range 800 °C to 2200 °C, at a number of national laboratories. An analysis of the calibration results confirms estimates of the accuracy of optical pyrometry in the range 800 °C to 2200 °C and supports the extension of these estimates to 4000 °C. The standard deviation uncertainty of optical pyrometry is shown to be about 1 °C at 800 °C rising to 2 °C at 2200 °C and 10 °C at 4000 °C, being about double this for the calibration of commercial pyrometers unless certain described precautions are taken.The reliability of the secondary standard lamps, when used under well-defined conditions, is confirmed and it is shown that they have a standard deviation calibration uncertainty of about 1 °C for the vacuum-type lamps in the range 800 °C to 1500 °C and 2 °C for the gas-filled lamps in the range 1500 °C to 2200 °C. Most of this uncertainty is due to primary standard optical pyrometer calibration errors. Attention is drawn to the fact that a carbon arc fulfills the requirements of a secondary luminance temperature standard at about 3514 °C.Recent determinations of the gold point and the second radiation constant indicate that the 1948 International Temperature Scale is lower than the thermodynamic scale by an amount varying from 0.8 °C at 800 °C to 12 °C at 4000 °C. This is already greater than the calibration errors of optical pyrometry and, in view of the still greater accuracies presaged by photomultipliers, a revision of the International Temperature Scale is suggested.
No takes yet. Share an insight, caveat, or question.
Derek Lovejoy (1958) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: