Key result
In-vitro testing of seven drug-eluting stent systems showed wide variations in mechanical properties, with pushability ranging from 18.53% to 38.66% and mean track-forces from 0.551 N to 1.137 N.
Why the study?
How do the mechanical performance characteristics compare among seven commercially available drug-eluting stent systems?
How do the mechanical performance characteristics compare among seven commercially available drug-eluting stent systems?
Commercially available drug-eluting stents exhibit a wide range of mechanical properties, highlighting the importance of understanding these characteristics for optimal stent selection during percutaneous coronary intervention.
Bench variations in DES deliverability may influence PCI strategy; leaves open correlation with procedural outcomes.
OBJECTIVES: Mechanical properties of drug eluting stents (DES) will be measured to provide comparable numerical data to assess deliverability, and thus clinical performance. BACKGROUND: DES are routinely used in coronary interventions to reduce the rates of restenosis and target vessel revascularizations. Current research is primarily concerned with issues related to late stent thrombosis. However, mechanical properties of DES are a critical determinant of deliverability, and consequently the ultimate arbiter of their clinical performance. METHODS: Mechanical properties (pushability, trackability, crossability) were measured under standardized in-vitro conditions. The vessel models were derived from typical vessel anatomy but adapted to the individual tests. Additionally, profile and bending forces of the stent segment of the delivery system were measured. Seven different commercially available balloon-expandable coronary DES systems were included. All stents were 3.0 mm diameter with a stent length from 14 to 18 mm. RESULTS: The pushability expressed as the ratio of distal force at a specific proximal push force (4N) ranged between 38.66 and 18.53%. The trackability as the mean track-forces ranged from 0.551 N to 1.137 N. One stent system could not pass this test. The mean crossing forces at a 1.4 mm stenosis model ranged from 0.038 N up to 0.103 N. The mean crimped stent profiles ranged from 1.055 mm to 1.198 mm and the bending stiffness of the crimped stent was 17.22 to 47.20 Nmm2. CONCLUSION: Better understanding of mechanical properties of DES shall improve tactile skills of the interventionists during PCI and to improve criteria for DES selection in specific clinical settings.
No takes yet. Share an insight, caveat, or question.
Schmidt et al. (2008) studied Coronary interventions. Drug-eluting stent systems was evaluated on Mechanical properties (pushability, trackability, crossability). In-vitro testing of seven drug-eluting stent systems showed wide variations in mechanical properties, with pushability ranging from 18.53% to 38.66% and mean track-forces from 0.551 N to 1.137 N.
Synapse has enriched 4 closely related papers on similar clinical questions. Consider them for comparative context: