This is the authors' abstract. We don't add key points for this paper.
Three designs of transurethral applicators have been analysed to find the one that is best able to selectively heat the desired volume of prostate. A helix-loaded-dipole-antenna (HLDA) inserted into a Foley type catheter was found to be the most promising design. A change in the heat deposition pattern of the antenna depending on prostate size is possible by moving the position of the antenna within a Foley catheter. A number of prototypes of HLDA were manufactured and tested to optimize their design. These tests were performed in solid and liquid phantoms and in laboratory animals. Intra-operative measurements of intra-prostatic temperature distribution were also performed. A comparison of the HLDA with other commercially available transurethral applicators and the published data showed that the present design has a higher linear homogeneity coefficient and a better heat deposition in the prostate.