Answers are grounded in this paper and cite their sources.
Characteristics of silicon and germanium radiation detectors are developed from an analysis of the energy-loss mechanisms for hot carriers. The carrier behavior is described in terms of phonon and ionization scattering rates. A calculation of the probability of ionization vs primary electron energy in silicon, using Kane's characterization of the scattering rates, indicates an effective ionization threshold several times the band-gap energy. Calculations of phonon losses and the carriers' residual kinetic energies are combined to give an average energy expended per created pair ε of 4.24 eV. Closer agreement with the experimental value of 3.67 eV is achieved by lowering the phonon to ionization scattering-rate ratio. Temperature effects, which enter through changes in the band gap, are calculated to be ∂ε/∂Eg=1.73. The analysis also predicts a Fano factor of 0.059 for silicon. Discussion of the germanium situation is based on the premise of hot-carrier behavior similar to that in silicon. Because germanium has a more complicated band structure, two sets of calculations are made using different effective band gaps. With ε=2.96 eV at 90 °K, very similar results are obtained in both cases and indicate ∂ε/∂Eg≈1.77 and F≈0.066.
No takes yet. Share an insight, caveat, or question.
Drummond et al. (1971) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: