Answers are grounded in this paper and cite their sources.
Electromechanical properties of thin films of vinylidene fluoride-trifluoroethylene copolymer [P(VDF-TrFE)] and their application to ultrasonic transducers were studied in the frequency range of 50–250 MHz at room temperature. A 5-μm-thick P(VDE-TrFE) film resonating at 250 MHz in thickness extensional mode exhibited a large electromechanical coupling factor of 0.26. The dielectric and elastic constants were also studied in this frequency region. A transducer using a 3-μm-thick film exhibited a large conversion efficiency (6.5 dB at 150 MHz), and good impulse response (30 ns ring-down time width) in water. A concave transducer transmitted a well-focused ultrasonic beam (35-μm beamwidth).