Answers are grounded in this paper and cite their sources.
The Micro-Media Systems (MMS) anaerobe susceptibility testing panel results from four laboratories were compared for interlaboratory and intralaboratory variations and for the results with the reference agar dilution and a broth microdilution method. The interlaboratory agreement was 98.0% and intralaboratory agreement was 97.3% (+/- 1 log2 dilution). When interpretive criteria for each antimicrobial agent (susceptible, intermediate, and resistant) were assigned, the MMS anaerobic minimum inhibitory concentration data showed an interpretive accuracy of 91.0 and 95.5% for comparisons to the reference agar dilution and the broth methods, respectively. Most significant interpretive errors were considered minor, and nearly half of all errors involved tetracycline, a drug rarely used for serious anaerobic infections. The MMS anaerobe panels appear to be acceptable for selected use in clinical microbiology laboratories.
No takes yet. Share an insight, caveat, or question.
Jones et al. (1982) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: