The Modulation Transfer Function (MTF) characteristics of radiochromic film (RCF) dosimetry with optical systems were evaluated with an RCF, HD-V2, which is transferred with the line patterns of the resolution test chart. In the evaluation using a GT-X980 flatbed scanner, a high contrast spatial resolution with an MTF of ≥0.8 is limited to about 83.3 μm due to artifacts such as scattering of the HD-V2 transmitted light. To achieve high contrast at higher spatial frequencies, a test bench microdensitometer (TBMD) was developed, and its optical performance, dose equivalent response, and spatial resolution characteristics with MTF were evaluated. The TBMD had a minimum readout focal diameter of 13 μm, and its high contrast spatial resolution with an MTF of ≥0.8 is about 11.9 μm. The result that the optical density error increases with decreasing focal diameter in the TBMD supports our hypothesis that these errors are based on the non-uniformity of the shading and surface roughness of the HD-V2 active layer monomers.