U isotopic ratio (UIR) in urine by inductively coupled plasma mass spectrometry (ICP-MS) using a Thermo Scientific iCAP-Q instrument. The performance characteristics of the assay were determined to be in compliance with clinical laboratory standards. The assay was linear in the concentration range of 1.0 to 500.0 ng/liter TU. The method was precise and accurate with limits of detection of 2.5 ng/liter for TU and 9.8 ng/liter for UIR. The accuracy was >93% and the coefficient of variation (% CV) was <5.0% for both TU and UIR. All results were within established guidelines and agreed-upon criteria, and the results fell within the certified range for the reference controls. The method has thus been shown to be effective as a simple, precise, and sensitive analytical technique for testing urine samples.