Answers are grounded in this paper and cite their sources.
An ultrahigh-voltage capillary electrophoresis system was built to demonstrate the possibility of extending the applied potential and thus the separation power of capillary electrophoresis. A commercial 30-kV power supply was extensively modified in order to provide electrical potentials up to 120 kV. A unique electrical shielding system was developed to prevent capillary breakdown and corona or spark discharges. Electrophoretic studies using a mixture of peptide standards, as well as a complex mixture of peptides obtained from a protein digest, showed that the numbers of theoretical plates achieved increase linearly with applied voltage. Theoretical plate counts ranging from 2.7 to 6.1 million plates were obtained for peptides in a separation done at 120 kV. Resolution also increased with the square root of applied voltage, as predicted by theory.
No takes yet. Share an insight, caveat, or question.
Hutterer et al. (1999) studied this question.