Answers are grounded in this paper and cite their sources.
Synthetic oxytocin (OT) conjugated to bovine thyroglobulin by the carbodiimide reaction was injected into rabbits to raise a high titre, specific OT antiserum which was then coupled to microcrystalline cellulose activated by cyanogen bromide. High affinity of the coupled antiserum was defined by Scatchard analysis, Keq = 7.1 X 10(11)1/mol. Cross-reactivity studies revealed little binding of antiserum to analogues of OT. Iodination was performed by the Chloramine T method, giving specific activity of 125I-OT, range 1.1 - 1.7 X 10(3) Ci/mmol. After incubation for 40 hours under disequilibrium conditions, specific and non-specific bindings were 10.6 +/- 2.7% and 0.2 +/- 0.1% (n=15), respectively. Displacement of 50% 125I-OT occurred with 2.9 pg OT/tube. Coefficients of variation of standard OT concentrations (0.03 - 16 pg/tube) were less than 5%. Limit of detection was 2 pg OT/ml plasma. Recovery of synthetic OT added to non-pregnant plasma was 81.8% (n=34) at 20 pg/ml and 97.4% (n=32) at 100 pg/ml. Two patients, 17 and 18 weeks post-partum, had increases in plasma OT from less than 2 pg/ml to 18.3 and 16.0 pg/ml after 6 and 4 minutes breast feeding infants, respectively. We conclude that this solid phase OT radioimmunoassay is quick, relatively sensitive and reliable, and does not require prior extraction of plasma samples.
No takes yet. Share an insight, caveat, or question.
Burd et al. (1985) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: