Answers are grounded in this paper and cite their sources.
An ultrashort-pulse Cr(2+):ZnSe laser is a novel broadband source for sensitive high resolution molecular spectroscopy. A 130-fs pulse allows covering of up to 380 cm(-1) spectral domain around 2.4 mum which is analyzed simultaneously with a 0.12 cm(-1) (3.6 GHz) resolution by a Fouriertransform spectrometer. Recorded in 13 s, from 70-cm length absorption around 4150 cm(-1), acetylene and ammonia spectra exhibit a 3800 signal-tonoise ratio and a 2.4.10(-7) cm(-1).Hz(-1/2) noise equivalent absorption coefficient at one second averaging per spectral element, suggesting a 0.2 ppbv detection level for HF molecule. With the widely practiced classical tungsten lamp source instead of the laser, identical spectra would have taken more than one hour.
No takes yet. Share an insight, caveat, or question.
Sorokin et al. (2007) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: