Objective . Clinical dosimetry in the presence of a 1.5 T magnetic field is challenging, let alone in case small fields are involved. The scope of this study is to determine a set of relevant correction factors for a variety of MR-compatible detectors with emphasis on small fields. Two dosimetry formalisms adopted from the literature are considered. Approach . Six small-cavity ionization chambers (from three manufacturers), four active solid-state detectors and a thermoluminescence dosimeter microcube were modeled in the EGSnrc Monte Carlo code. Phase space files for field sizes down to 1 × 1 cm 2 of the Unity 1.5 T/7 MV MR-linac (Elekta, UK) were used as source models. Simulations were performed to calculate the k Q B , Q f B , f (also known as k B , Q ), k Q msr B , f msr and k Q clin , Q msr B , f clin , f msr relevant to two different dosimetry formalisms. Two detector orientations with respect to the magnetic field were considered. Moreover, the effect of the ionization chamber’s stem length (a construction parameter) on the correction factor was investigated. Simulations were also carried out to determine whether correction factors obtained in water can be applied in dosimetry procedures involving water-equivalent solid phantoms. Main results . Under the k Q B , Q f B , f -based formalism, the required corrections to ionization chamber responses did not exceed 1.5% even for the smallest field size considered. A much wider range of k Q