Answers are grounded in this paper and cite their sources.
The routine application of capillary electrochromatography (CEC) is demonstrated by incorporating 75 microns I.D. capillaries packed with 3 microns octadecylsilica (ODS) particles into a commercial CZE instrument. A mixture of several neutral compounds is separated into its components with an average efficiency up to 181 000 plates/m in less than 8 min. Hundreds of consecutive runs are performed over a period of weeks from which it is concluded that the reproducibility of the capacity factors is better than 2% and that CEC separations can be achieved in a reliable and routine manner.
No takes yet. Share an insight, caveat, or question.
Dulay et al. (1996) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: