Answers are grounded in this paper and cite their sources.
Calibrations of commercially available quartz-coated and lacquer-coated conical hot platinum film probes indicate a downward shift in sensitivity due to film deterioration, a sudden increase in cold resistance preceding catastrophic failure, a radial sensitivity one third greater than the axial sensitivity, ambiguities in the value of the exponent n in King's law, and a maximum operating life of 20.5 hours. Field experience indicates poor reliability and an inordinate number of required experimental precautions.
No takes yet. Share an insight, caveat, or question.
Frey et al. (1973) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: