Experimental study demonstrates an all-optical microfluidic platform for single-cell protein quantification in breast cancer cells, highlighting a viable route toward single-cell proteomics.
We have developed a generic platform to undertake the analysis of protein copy number from single cells. The approach described here is 'all-optical' whereby single cells are manipulated into separate analysis chambers using an optical trap; single cells are lysed by a shock wave caused by laser-induced microcavitation, and the protein released from a single cell is measured by total internal reflection microscopy as it is bound to micro-printed antibody spots within the device. The platform was tested using GFP transfected cells and the relative precision of the measurement method was determined to be 88%. Single cell measurements were also made on a breast cancer cell line to measure the relative levels of unlabelled human tumour suppressor protein p53 using a chip incorporating an antibody sandwich assay format. These results suggest that this is a viable method for measuring relative protein levels in single cells.
No takes yet. Share an insight, caveat, or question.
Salehi‐Reyhani et al. (2011) studied this question.