This is the authors' abstract. We don't add key points for this paper.
Thirty‐four women with type III stress urinary incontinence underwent implantation of the artificial urinary sphincter (AUS) for control of urination using the AMS #800 TM device via a combined vaginal and minimal retropubic approach. The deliberate opening of the vagina facilitates placement of the sphincteric cuff in its proper position and serves to prevent urethral or bladder neck damage without increasing the risks of cuff infection or erosion. Results show no erosions, no detrusor instability, four patients on intermittent self‐catheterization, and three patients requiring repeat surgery for mechanical problems. These results indicate that the AUS is a viable alternative in the management of this difficult group of incontinent women and that implanting surgeons should not fear entering the vagina during implantation, but should deliberately do so to facilitate surgical placement of the cuff while protecting the urinary organs and reducing operating room time.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Rodney A. Appell (1988) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: