13 C breath testing is increasingly used in physiology and ecology research because of what it reveals about the different fuels that animals oxidize to meet their energetic demands. Here I review the practice of 13 C breath testing in humans and other animals and describe the impact that contamination by ambient/background CO 2 in the air can have on the accuracy of 13 C breath measurements. I briefly discuss physical methods to avoid sample contamination as well as the Keeling plot approach that researchers have been using for the past two decades to estimate δ 13 C from breath samples mixed with ambient CO 2 . Unfortunately, Keeling plots are not suited for 13 C breath testing in common situations where (1) a subject's VCO 2 is dynamic, (2) ambient [CO 2 ] may change, (3) a subject is sensitive to hypercapnia, or (4) in any flow‐through indirect calorimetry system. As such, I present a mathematical solution that addresses these issues by using information about the instantaneous [CO 2 ] and the δ 13 CO 2 of ambient air as well as the diluted breath sample to back‐calculate the δ 13 CO 2 in the CO 2 exhaled by the animal. I validate this approach by titrating a sample of 13 C‐enriched gas into an air stream and demonstrate its ability to provide accurate values across a wide range of breath and air mixtures. This approach allows researchers to instantaneously calculate the δ 13 C of exhaled gas of humans or other animals in real time without having to scrub ambient CO 2 or rely on estimated values.