Answers are grounded in this paper and cite their sources.
High-resolution capillary electrophoresis has been coupled to MALDI-TOF and TOF/TOF MS through off-line vacuum deposition onto standard stainless steel MALDI targets. This off-line approach allowed the decoupling of the separation from the MS analysis, thus allowing each to be independently optimized in terms of time. Using BSA tryptic digest as a model sample, the deposited streaks, roughly 100-microm wide, were first analyzed in the MS mode, consuming only a fraction of the sample. After data analysis, segments of the deposited trace, containing unidentified peptides, as well as several species chosen for sequence confirmation, were reanalyzed in the MS/MS mode using MALDI-TOF/TOF MS. Additionally, it is shown that the shot-to-shot reproducibility of the vacuum-deposited trace (5% RSD) is 1 order of magnitude lower than that found for the standard dried droplet method. Moreover, a linear dependence of signal intensities (relative to an internal standard) over 3 orders of magnitude was found for a peptide sample with concentrations ranging from 1 to 1000 nM. This paper demonstrates the potential of off-line coupling of high-resolution separations to MALDI-MS and MALDI-MS/MS using vacuum deposition for the analysis of complex peptide mixtures from protein digests.
No takes yet. Share an insight, caveat, or question.
Rejtar et al. (2002) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: