A new device for continuous mechanical introduction of liquid samples into a mass spectrometer is described. A ball mounted on a shaft is supported by a polymer gasket at the circumference of a hole leading into the mass spectrometer. The surface of the ball on the outside is in contact with the liquid sample. Sample sticking to the surface of the ball is dragged past the gasket into the mass spectrometer where it evaporates and is ionized by electron impact. The rotating ball inlet overcomes some of the problems associated with previously known methods for continuous sample introduction into a mass spectrometer. A wide variety of compounds can be measured continuously in aqueous solution with a response time in the order of seconds. The fragmentation of molecules caused by electron impact ionization facilitates identification of analytes.