Answers are grounded in this paper and cite their sources.
A bead-bed immunoassay system was structured on a microchip and applied to determine carcinoembryonic antigen (CEA), which is a commonly used marker of colon cancer. Polystyrene beads precoated with anti-CEA antibody were introduced into a microchannel, and then a serum sample containing CEA, the first antibody, and the second antibody conjugated with colloidal gold were reacted successively. The resulting antigen-antibodies complex, fixed on the bead surface, was detected using a thermal lens microscope (TLM). A highly selective and sensitive determination of an ultratrace amount of CEA in human sera was made possible by a sandwich immunoassay system that needs three antibodies for an assay. A detection limit dozens of times lower than the conventional ELISA was achieved. Moreover, when serum samples for 13 patients were assayed with this system, there was a high correlation (r = 0.917) with the conventional ELISA. The integration reduced the time necessary for the antigen-antibody reaction to approximately 1%, thus shortening the overall analysis time from 45 h to 35 min. Moreover, troublesome operations required for conventional heterogeneous immunoassays could be much simplified. This microchip-based diagnosis system is the first microchip-based system that is practically useful for clinical diagnoses with short analysis time, high sensitivity, and easy procedures.
No takes yet. Share an insight, caveat, or question.
Sato et al. (2001) studied this question.