We have evaluated three anti-hepatitis E virus (anti-HEV) immunoglobulin M (IgM) assays, the EIAgen HEV IgM assay (Adaltis), the HEV IgM enzyme-linked immunosorbent assay 3.0, and the Assure HEV IgM rapid test (MP Diagnostics), for the routine detection of acute genotype 3 HEV. Their sensitivities were fairly good (90%, 88%, and 82%), and their specificities were excellent (100%, 99.5%, and 100%).
No takes yet. Share an insight, caveat, or question.
Abravanel et al. (2009) studied this question.