Why the study?
Does a closed-loop vagus nerve stimulation system based on state transition models improve heart rate regulation compared to a proportional-integral controller in sheep?
Does a closed-loop vagus nerve stimulation system based on state transition models improve heart rate regulation compared to a proportional-integral controller in sheep?
A novel closed-loop vagus nerve stimulation control system based on state transition models provides satisfactory accuracy and time response for heart rate regulation in a sheep model.
Satisfactory metrics in sheep; leaves open translation of STM-based closed-loop VNS to human HR regulation.
OBJECTIVE: Vagus nerve stimulation (VNS) is a potential therapeutic approach in a number of clinical applications. Although VNS is commonly delivered in an open-loop approach, it is now recognized that closed-loop stimulation may be necessary to optimize the therapy. In this paper, we propose an original generic closed-loop control system that can be readily integrated into an implantable device and allows for the adaptive modulation of multiple VNS parameters. METHODS: The proposed control method consists of a state transition model (STM), in which each state represents a set of VNS parameters, and a state transition algorithm that optimally selects the best STM state, minimizing the error between an observed physiological variable and a given target value. The proposed method has been integrated into a real-time adaptive VNS prototype system and has been applied here to the regulation of the instantaneous heart rate, working synchronously with cardiac cycles. A quantitative performance evaluation is performed on seven sheep by computing classical control performance indicators. A comparison with a proportional-integral (PI) controller is also performed. RESULTS: The STM controller presents a median mean square error, overshoot, and settling time, respectively, equal to 622.21 ms , 72.8%, and 7.5 beats. CONCLUSION: The proposed control method yields satisfactory accuracy and time response, while presenting a number of benefits over classical PI controllers. It represents a feasible approach for multiparametric VNS control on implantable devices. SIGNIFICANCE: Closed-loop multiparametric stimulation may improve response and minimize side effects on current pathologies treated by VNS.
No takes yet. Share an insight, caveat, or question.
Ugalde et al. (2017) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: