Answers are grounded in this paper and cite their sources.
PURPOSE: Core temperature (Tcore) monitoring is used in the prevention of heat illnesses and for heat-acclimation purposes. We examined the accuracy and precision of 2 commercially available devices (BandV2 and CORE) that estimate Tcore versus rectal temperature. METHOD: Eight eumenorrheic females (V˙O2max: ∼41 mL·kg-1·min-1) completed 60 minutes of cycling in the follicular phase and the luteal phase over 2 separate cycles, wearing a minimally permeable clothing ensemble to amplify thermal load. RESULTS: Both devices proved to be precise at rest and during exercise. Between duplicate follicular and luteal tests, the CORE device bias was 0.1400 (0.33) °C and 0.0331 (0.42) °C, and the BandV2 device bias was 0.0418 (0.18) °C and -0.0171 (0.21) °C. Compared with rectal temperature, accuracy was below our preestablished criterion of ±0.27 °C. At rest, the devices underestimated Tcore: BandV2, -0.2735 (0.25) °C, and CORE, -0.2746 (0.28) °C, and at the 55-minute time point, both devices overestimated Tcore: BandV2, +0.5117 (0.37) °C, and CORE, +0.3319 (0.43) °C. The delta increase in Tcore did not differ between menstrual-cycle phases. CONCLUSIONS: The BandV2 and CORE indirect sensors currently offer precise but not accurate estimates of Tcore.
No takes yet. Share an insight, caveat, or question.
Richard et al. (2025) studied this question.