Answers are grounded in this paper and cite their sources.
Objective. To develop a simple and rapid high-performance liquid chromatography (HPLC) method for measuring moxifloxacin concentration in human plasma. Methods. Following a single step liquid-liquid extraction, analytes along with an internal standard (IS) were separated using an isocratic mobile phase of 0.1% triethylamine (adjusted pH to 4.8 with phosphoric acid)/acetonitrile (80/20, v/v) at flow rate of 1 mL/min on reverse phase Kromasil C18 column (250 mm × 4.6 mm, 5 μ m) at room temperature. Results. Total analytical run time for selecting moxifloxacin was 15 min. The assays exhibited good linearity (r (2) = 0.9998) over the studied range of 25 to 5000 ng/mL. The absolute recovery rate of low, medium, and high concentrations were 69.88%, 78.86%, and 78.51%, respectively. The relative recovery rates were 98.50%, 96.61%, and 101.79%, respectively. Coefficient of variation and error at both of the intraday and interday assessments were less than 4.7%. Conclusions. The results indicated that this method is a simple, rapid, precise and accurate assay for the determination of moxifloxacin concentrations in human plasma. This validated method is sensitive and reproducible enough to be used in pharmacokinetic studies.
No takes yet. Share an insight, caveat, or question.
Wang et al. (2013) studied this question.