Key result
An approximate threshold for Korotkoff sound audibility was obtained by comparing signal-to-noise ratios of sounds to their masking counterparts with difference limen in psychoacoustic literature.
Why the study?
While absolute sound level audibility is well researched, Korotkoff sound audibility during blood pressure measurement had not been evaluated from the perspective of psychoacoustic masking.
Applying psychoacoustic masking principles to Korotkoff sounds provides an objective threshold for audibility, which could improve automated blood pressure measurement devices.
Measurement of blood pressure (BP) through manual auscultation and the observation of Korotkoff sounds (KSs) remains the gold standard in BP methodology. Critical to determining BP levels via auscultation is the determination of KS audibility. While absolute sound level audibility is well researched, the problem has not been approached from the point of view of psychoacoustic masking of the sounds. Here, during manual auscultation of BP, a direct comparison is made between what an observer perceives as audible and the electronic analysis of audibility level determined from masking of sound signal levels. KSs are collected during auscultation with an electronic stethoscope, which allows simultaneously observing sound audibility and recording the sound electronically. By time-segmenting the recorded sound around Korotkoff peaks into a test segment and a masking segment, performing Fourier transforms on the segments, and comparing frequency-band sound energy levels, signal-to-noise ratios of a sound to its masking counterpart can be defined. Comparing these ratios to difference limen in the psychoacoustic masking literature, an approximate threshold for sound audibility is obtained. It is anticipated that this approach could have profound effects on future development of automated auscultation BP measurements.
No takes yet. Share an insight, caveat, or question.
Alvarez et al. (2023) studied Blood pressure measurement. Electronic analysis of audibility level using psychoacoustic masking vs. Observer perception of audibility (manual auscultation) was evaluated on Threshold for sound audibility. An approximate threshold for Korotkoff sound audibility was obtained by comparing signal-to-noise ratios of sounds to their masking counterparts with difference limen in psychoacoustic literature.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: