Key result
Intracardiac echocardiography guidance for balloon mitral valvuloplasty was feasible, with mean mitral valve area increasing from 0.9 to 1.7 cm2 (P<0.0001).
Why the study?
Is intracardiac echocardiography (ICE) guidance feasible and useful during balloon mitral valvuloplasty (BMV) in patients with mitral stenosis?
Observational (n=17)
Is intracardiac echocardiography (ICE) guidance feasible and useful during balloon mitral valvuloplasty (BMV) in patients with mitral stenosis?
p-value: p=<0.0001
ICE guidance during balloon mitral valvuloplasty is feasible and provides comparable hemodynamic measurements to TTE and catheterization while aiding in procedural guidance and complication detection.
BACKGROUND AND AIMS: BMV is an established treatment for rheumatic mitral valve stenosis. The procedure is historically guided by fluoroscopy, and the role of intracardiac echocardiogram (ICE) guidance is not well defined. We report our initial experience of using ICE to guide BMV procedures. METHODS: During BMV procedure, ICE catheter was inserted into the right atrium from the right femoral vein, and the septal puncture was monitored by ICE, as well as positioning of the balloon in the mitral valve. Comparisons were made between ICE, transthoracic echocardiography (TTE), and catheterization derived hemodynamic measurements (cath). RESULTS: Seventeen patients with mitral stenosis underwent the procedure. The mean age was 44.4 ± 21 years. The mean MV area increased from 0.9 ± 0.1 cm(2) to 1.7 ± 0.2 cm(2), P < 0.0001 and the mean gradient decreased from 12.6 ± 5.8 mmHg to 4.9 ± 1.8 mmHg, P < 0.001. Atrial septum puncture and guidance of the balloon into the MV apparatus were obtained in all patients under ICE guidance. Severe MR developed in one patient and was readily detected by ICE. ICE derived gradient measurements were comparable to those obtained by TTE, and cath. CONCLUSION: ICE guidance of BMV is feasible, and useful in monitoring safe septal puncture, optimizing balloon positioning, and in detecting complications. The hemodynamic measurements obtained were comparable to those obtained by TTE, and cath.
No takes yet. Share an insight, caveat, or question.
Ahmari et al. (2011) conducted an observational in Mitral stenosis (n=17). Intracardiac echocardiography (ICE) guidance vs. Transthoracic echocardiography (TTE) and catheterization was evaluated on Mean mitral valve area and mean gradient (p=<0.0001). Intracardiac echocardiography guidance for balloon mitral valvuloplasty was feasible, with mean mitral valve area increasing from 0.9 to 1.7 cm2 (P<0.0001).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: