The objective of this study was to evaluate changes in bone density in the distal femur 2 years after total knee arthroplasty with four different implant designs using cemented or cementless femoral components. Bone density was measured retrospectively from radiographs of 114 knees, using a densitometer. A decrease in bone density of up to 57% was identified in the distal femora with a cemented femoral component 2 years after surgery, compared with a decrease of up to 28% with a cementless, porous-coated component of the same design. The differences between the four implant designs in the changes in bone density in the anterior distal femur were significant at 2 years (P < 0.001). A possible cause of this result may be a difference in load transfer due to the different methods of fixation. The surgeon should expect decreased bone density in the distal femur at the time of revision surgery, especially with a cemented femoral component.