A hand-held dissolved oxygen optical sensor based on solid-state electronics and highly oxygen-sensitive luminescence material has been developed. Oxygen-sensitive dye absorbed on silica gel particles was dispersed in a 0.2 mm homogenous silicone rubber film (optode membrane) and coated on a 580 nm long-pass filter. The O 2 -sensitive dye was excited by an ultra-bright blue light-emitting diode and the emission intensity was captured by a silicon photodiode. The long-pass filter efficiently reduced the scattered light from the light source. The emission signal of the optode membrane was converted to the voltage signal which was processed by a single-chip Micyoco controller, and the dissolved oxygen concentration was automatically displayed on a liquid crystal display unit. The developed hand-held optical sensor is a single-piece device which can provide instant and reliable dissolved oxygen measurement. The sensor showed good repeatability, photostability and long-term stability. Satisfactory results were obtained when its analytical performance was compared with that of a typical Clark-type amperometric oxygen electrode. The developed oxygen optical sensor possesses the advantages of portability, low cost and long lifetime.
No takes yet. Share an insight, caveat, or question.
Xiao et al. (2003) studied this question.