Answers are grounded in this paper and cite their sources.
A general method of preparation of thin-film sensors for O 2 , incorporating the dye ion-pair tris(4,7-diphenyl-1,10-phenanthroline) ruthenium(II) ditetraphenylborate, in a variety of different thin film polymer/plasticizer matrices is described. The sensitivity of the sensor depends upon the nature of the polymer matrix and plasticizer. A detailed study of one of these systems utilising the polymer poly(methyl methacrylate), PMMA, is reported. The sensitivity of this O 2 sensor depends markedly upon the plasticizer concentration and is largely independent of temperature (24.5–52.5 °C) and age (up to 30 d). When exposed to an alternating atmosphere of O 2 and N 2 , a typical oxygen film sensor in PMMA exhibits a 0–90% response and recovery time of 0.4 and 4.5 s, respectively.
No takes yet. Share an insight, caveat, or question.
Mills et al. (1997) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: