Answers are grounded in this paper and cite their sources.
A hot stage temperature controller with automatic transition between linear sweep and constant temperature modes, using only two operational amplifiers, is described. The derivative of temperature is nulled against a fixed reference signal instead of the usual approach in which temperature is compared with a time−dependent reference. This allows an increased loop gain at high speeds, while allowing the Ṫ−T crossover to be achieved without additional amplifiers. Ramp speeds of 0.1−10 K/sec in the range 300−820 K are obtained with accuracy of 1%−3%. Noise− and transient−free dc heater power simplifies shielding requirements in the 10−15 A range. A new approximation of the type K thermocouple characteristic is used to estimate its influence on linearity.
No takes yet. Share an insight, caveat, or question.
William L. Paterson (1975) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: